M082 series super-high vacuum hot cathode ionization gauge, M082, M082A, Ionization gauge, EECTECH
M082 is equivalent to American Varian company 571, G-P company G-75-P, Veeco company RG-75, etc.
M082A is equivalent to American Varian Company 572, Veeco Company TG-75, etc.
Introduce:
1. Although M082 and M082A hot cathode ultra-high vacuum ionization gauges are different in structure, they are quite similar in terms of technical and working parameters.
2. M082 can be interchanged with American Varian Company 571, G-P Company G-75-P, Veeco Company RG-75, etc.
3. M082A can be interchanged with American Varian 572, Veeco TG-75 and other regulations.
4. The difference is that M082 type ionization gauge uses electrophoretic yttrium oxide filament, which has strong oxidation resistance and long life; M082A uses spiral double W filament (one spare), which also compensates for the short life of W filament to a certain extent.
5.Users can choose the appropriate flange interface according to their own system requirements. Because there are many models, we only show typical models and accept customization.
Technical parameters:
Measuring Range: 1X10-1---4X10-2Pa
X Ray limit: 4X10-4Pa
Gauge sensitivity: 0.075Pa
Above earth potential of accelerating electrod: 180V
Above earth potential of cathode: 30V
Above earth potential of collectrode: 0V
Emission current: 1mA
Electron bombardment exhaust: Max power to be 75W
Delivery time:
Usually, we will ship within 7-10 days after order confirmation. and then we will try to ship as soon as possible, or we will contact you with suggestions.
Bulk order:
If you want to order in bulk, please contact us for better prices and better shipping costs.